ANALYZING THE CONSISTENCY OF BEAM QUALITY AND OUTPUT DOSE/MU FOR PHOTON BEAMS OF ENERGY 6 MV AND 15 MV OF A MEDICAL LINAC USED FOR RADIOTHERAPY AT INSTITUTE OF NUCLEAR MEDICAL PHYSICS, AERE, SAVAR.
Abstract & Details
Research Area
Radio Therapy, Medical Physics
Keywords
Dosimetry
Consistency
Beam Quality
Output Dose
Abstract
Medical Linac (linear accelerator) delivers ionizing radiation which is used for cancer management. In radiotherapy, it is mandatory to provide maximum dose to the cancerous tumor and minimum dose to the surrounding normal tissues. So it is required to assess the consistency of beam quality and beam output dose/MU (Monitor Unit).
The output dose of Linac (Model: CLINAC-iX) of photon beam of energy 6 MV and 15 MV installed at Institute of Nuclear Medical Physics, AERE Savar has been measured with FC65-P ionization chamber using 1D Water Phantom. In this study, data collection period was from October 2019 to September 2022 in accordance with the TRS-398 dosimetry protocols.
Over the period of 36 months we found that TPR20,10 for 6 MV photon beam was 0.67 corresponding and TPR20,10 for 15 MV photon was 0.76. Which showed that the values of beam quality TPR20,10 never changed when the values are rounded to two decimal points. The measured output dose/MU was found in between 0.998 to 1.023 cGy for 6MV photon and 0.998 to 1.032 cGy for 15MV photon. The dose/MU for all radiation beams was maintained within 2% of deviation from the recommended value 1cGy/MU at Dmax position over the period of 36 months. In our study, the observed deviations of output dose/MU were maintained in tolerance interval instructed by the protocol. If the tolerance crossed the machine tuned of the monitor unit for getting 1 cGy/MU at the Dmax position in water. Here, absorbed dose to water based protocol was followed which reduced the uncertainty of the accuracy of the dose determination.
License
This work is licensed under a Creative
Commons
Attribution-ShareAlike 4.0 International License.
Author Information
| # | Name | Institute / Affiliation |
|---|---|---|
| 1 | M.A. Hasnat | Institute of Nuclear Medical Physics |
| 2 | K.M.M. Morshed | Institute of Nuclear Science and Technology |
| 3 | M.M. Parvej | Institute of Nuclear Medical Physics |
| 4 | M.S. Sultana | Institute of Nuclear Medical Physics |
| 5 | Juwel Hosen | Institute of Nuclear Medical Physics |
| 6 | M.M. Ahasan | Institute of Nuclear Medical Physics |
| 7 | Shirin Akter | Department of Health Physics, Atomic Energy Centre |
| 8 | A.N. Monika | Department of Health Physics, Atomic Energy Centre |
How to Cite
Use the following formats to cite this article in your research.
APA Style
Hasnat, M.A., Morshed, K.M.M., Parvej, M.M., Sultana, M.S., Hosen, Juwel, Ahasan, M.M., Akter, Shirin, & Monika, A.N. (2022). ANALYZING THE CONSISTENCY OF BEAM QUALITY AND OUTPUT DOSE/MU FOR PHOTON BEAMS OF ENERGY 6 MV AND 15 MV OF A MEDICAL LINAC USED FOR RADIOTHERAPY AT INSTITUTE OF NUCLEAR MEDICAL PHYSICS, AERE, SAVAR.. International Journal of Advance Research and Innovative Ideas In Education, 8(5), 898-903.
MLA Style
Hasnat, M.A., et al. "ANALYZING THE CONSISTENCY OF BEAM QUALITY AND OUTPUT DOSE/MU FOR PHOTON BEAMS OF ENERGY 6 MV AND 15 MV OF A MEDICAL LINAC USED FOR RADIOTHERAPY AT INSTITUTE OF NUCLEAR MEDICAL PHYSICS, AERE, SAVAR.." International Journal of Advance Research and Innovative Ideas In Education, vol. 8, no. 5, 2022, pp. 898-903.
IEEE Style
M.A. Hasnat, K.M.M. Morshed, M.M. Parvej, M.S. Sultana, Juwel Hosen, M.M. Ahasan, Shirin Akter, and A.N. Monika, "ANALYZING THE CONSISTENCY OF BEAM QUALITY AND OUTPUT DOSE/MU FOR PHOTON BEAMS OF ENERGY 6 MV AND 15 MV OF A MEDICAL LINAC USED FOR RADIOTHERAPY AT INSTITUTE OF NUCLEAR MEDICAL PHYSICS, AERE, SAVAR.," International Journal of Advance Research and Innovative Ideas In Education, vol. 8, no. 5, pp. 898-903, 2022.
Vancouver Style
Hasnat M.A., Morshed K.M.M., Parvej M.M., Sultana M.S., Hosen Juwel, Ahasan M.M., et al. ANALYZING THE CONSISTENCY OF BEAM QUALITY AND OUTPUT DOSE/MU FOR PHOTON BEAMS OF ENERGY 6 MV AND 15 MV OF A MEDICAL LINAC USED FOR RADIOTHERAPY AT INSTITUTE OF NUCLEAR MEDICAL PHYSICS, AERE, SAVAR.. International Journal of Advance Research and Innovative Ideas In Education. 2022;8(5):898-903.
Harvard Style
Hasnat, M.A., Morshed, K.M.M., Parvej, M.M., Sultana, M.S., Hosen, Juwel, Ahasan, M.M., Akter, Shirin, & Monika, A.N. (2022) 'ANALYZING THE CONSISTENCY OF BEAM QUALITY AND OUTPUT DOSE/MU FOR PHOTON BEAMS OF ENERGY 6 MV AND 15 MV OF A MEDICAL LINAC USED FOR RADIOTHERAPY AT INSTITUTE OF NUCLEAR MEDICAL PHYSICS, AERE, SAVAR.', International Journal of Advance Research and Innovative Ideas In Education, 8(5), pp. 898-903.
Chicago Style
Hasnat, M.A., et al. "ANALYZING THE CONSISTENCY OF BEAM QUALITY AND OUTPUT DOSE/MU FOR PHOTON BEAMS OF ENERGY 6 MV AND 15 MV OF A MEDICAL LINAC USED FOR RADIOTHERAPY AT INSTITUTE OF NUCLEAR MEDICAL PHYSICS, AERE, SAVAR.." International Journal of Advance Research and Innovative Ideas In Education 8, no. 5 (2022): 898-903.
Turabian Style
Hasnat, M.A., et al. "ANALYZING THE CONSISTENCY OF BEAM QUALITY AND OUTPUT DOSE/MU FOR PHOTON BEAMS OF ENERGY 6 MV AND 15 MV OF A MEDICAL LINAC USED FOR RADIOTHERAPY AT INSTITUTE OF NUCLEAR MEDICAL PHYSICS, AERE, SAVAR.." International Journal of Advance Research and Innovative Ideas In Education 8, no. 5 (2022): 898-903.
Related Research
A Comprehensive Review of Non-Conventional Energy Sources: Applications, Challenges, and Global Prospects
PDF Unavailable
4G Model of Heavy Electroweak Charged 585 GeV Fermions as the Supposed Microscopic Origin of the 1.17 TeV All-Electron Spectral Break
PDF Unavailable
“Refractive index as a function of solute concentration: A case study with sugar solutions”
PDF Unavailable
Teaching and learning process to enhance teaching effectiveness
PDF Unavailable
Emerging Trends in Renewable Energy: Innovations and Future Directions for Solar, Wind, and Geothermal Power
PDF Unavailable
THE FUTURE OF SPACE EXPLORATION: PHYSICS CHALLENGES AHEAD
PDF Unavailable